Significant decline in galactomannan signal during storage of clinical serum samples

Download (321.75 kB)
journal contribution
posted on 2023-07-26, 13:14 authored by Gemma L. Johnson, Shah-Jalal Sarker, Kate Hill, Dimitris A. Tsitsikas, Amelie Morin, Stephen A. Bustin, Samir G. Agrawal
Galactomannan (GM) is widely used for detection of invasive aspergillosis in high-risk haemato-oncology patients. Recent publications have reported a lack of repeatability of GM detection. The objective of this retrospective study was to assess the repeatability of GM levels during storage of clinical samples. In a GM screening strategy, positive sera were repeat tested as per manufacturer's recommendations. Short-term (ST) storage of samples was at +4 °C while long-term (LT) storage was at -80 °C. Bronchoalveolar (BAL) fluid was also repeating tested after ST storage and LT storage. Wilcoxon Signed Ranks Test was employed to assess the repeatability of GM levels. In a subset of 14 GM positive sera, repeat testing was performed on both the original serum and ethylenediaminetetraacetic acid (EDTA) pre-treated sample. There was a significant reduction in GM signals on repeat testing following ST storage (median GM index: 0.65 vs. 0.19; p < 0.001) and LT storage (median GM index: 0.56 vs. 0.10; p < 0.001) of serum samples. Of samples that were initially GM positive, an average GM index reduction of 50% was seen, with approximately two-thirds becoming GM negative on repeat testing of the same sample. In contrast, GM signal loss was not seen on repeat testing of BAL fluid following ST or LT storage. When GM positive serum samples were repeat tested using EDTA pre-treated serum from the first step of the testing protocol, all samples remained GM positive. In contrast, when the same samples were repeat tested from the original collected serum, 9 samples (64%) became GM negative. The significant reduction in GM signals during ST and LT storage of serum samples has implications for clinical management. Although the reasons for GM decline are unknown, they occur prior to the EDTA pre-treatment stage, indicating that the time from phlebotomy to testing should be minimized. BAL fluid GM index values remain stable.

History

Refereed

  • Yes

Volume

14

Issue number

7

Page range

12970-12977

Publication title

International Journal of Molecular Sciences

ISSN

1422-0067

Publisher

MDPI

File version

  • Published version

Language

  • eng

Legacy posted date

2013-07-08

Legacy creation date

2019-08-01

Legacy Faculty/School/Department

ARCHIVED Faculty of Health, Social Care & Education (until September 2018)

Usage metrics

    ARU Outputs

    Categories

    No categories selected

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC